*Detector: Created a www-page which shows some of the known CCD3-controller features: the oscillation of the bias level, vertical strips seen in xbinned images and over-exposing in binned mode.
Observes noted that when using binning with ALFOSC you get additional
noise in the form of vertical lines. It was found that the vertical
strips seen in xbinned images is due to the lose of synchronisation of
the controller power supply with the pixel readout timing. When you do
binning in X the pixel time increases so giving a none integer period
for the PSU to synchronise. For xbin 2 reducing the 'tsam' from 172
clock to 136 clock returns the synchronisation, and no strips are
seen. The shorter sample time does increase the noise by 1 ADU
(0.3 e-) compared with no binning.
The Dewar Interface Board (DIB) in the CCD3 controller failed again
with identical symptoms as before, a bias level of 9100 ADU and
no signal from amplifier A and a very high level
700k ADU from
amplifier B. The board was replaced and two new boards were requested
(and received) from NBI, Copenhagen, Denmark.
Thomas Augusteijn 2016-05-05