StanCam and MOSCA together
StanCam is in principle always prepared when MOSCA is mounted, but if you
plan to use it, please, state so explicitly in your
NIRF . StanCam is per default equipped
with standard UBVRIz and H-alpha filters.
This is a quick reference for using StanCam and MOSCA together,
see details from the
StanCam cookbook
- Open another desktop on the lisa terminal where you are already logged
in on marissa and running MOSCA. Open an xterm window and login on tessa
with ssh -l obs tessa, the password being provided by staff.
-
In the tessa xterm type startobssys stancam.
A window will pop up asking for the TCS access code. This code is found on the
TCS in status page 1 which you get by typing sh 1 on the TCS command line.
The code is shown on the top part of the screen as: Access Code xxxxxxxxxxx.
Type it in. Note that you have to do this every night, as with all instruments.
- To save current TCS settings
for MOSCA, if needed, type in the Sequencer window
tcs.store-tel <filename> .
The data file will be stored in the current working directory.
If the file already exists, it will be overwritten.
- In addition to the 3 StanCam BIAS windows, also a
Sequencer window and a Talker will be
started up. If the talker is already running for MOSCA on the other desktop,
it will not be visible on this desktop. (The plan is to have this window
sticky on all desktops by default.)
- To setup the TCS correctly for StanCam imaging, in the Sequencer window type: tcs.setup-tel-stancam
If you wish to know the content of this script, it is explained
here.
Please, note that the default telescope focus value for StanCam (23040) is
used. (Telescope power should be on when you run this script.)
- Select the StanCam filter from the sequencer window by typing
ccd-filter or from the TCS
CCD-fil and the number corresponding to
the filter you want (U=1,B=2,V=3,R=4,I=5,Ha=6, SDSSz=7), e.g.
[obs@tessa]$ ccd-filter 4 (sequencer window)
or CCD-fil 4 (TCS command line).
- Focus the telescope and the guide camera:
- If default focus is good enough then use foc-del on the TCS.
The focus offset between ALFOSC (no filter) and the StanCam R-band
filter is -260 units, in the sense that going from MOSCA to StanCam you
apply foc-del -100.
- If the optimum StanCam focus is required
then use a focus star close to zenith (e.g. from
the blanks catalog) and start a focus sequence with the BIAS command
focus (start value) (focus step-size) (exptime)
(e.g. focus 22920 80 12) **do not autoguide**. The resulting image
gives a number of exposures at different focus values (the last one is offset
by two steps), and you can evaluate which is the best focus
(see focussing StanCam).
If necessary,
iterate.
- Finally, correct the focus on the guide TV if autoguiding is needed.
Start with
tv-foc 415 and adjust with steps of 10.
- Make an exposure by typing expose followed by the time in seconds
(see
STANCAM SEQUENCER Command Reference).
Take single target exposures or run your scripts.
- When finished with StanCam to restore the MOSCA TCS settings
type in the Sequencer window
tcs.restore-tel <filename> .
This command restores saved TCS settings, stored by tcs.store-tel <filename>
The data will be read from file in the current working directory.
- When you want to exit from the StanCam Sequencer type
shutdownobssys stancam.
Comments to Tapio Pursimo
|