Spectroscopy with NOTCAM
Near-IR spectroscopy
NOTCAM can be used as an imager and as a spectrograph. The switch
between imaging and spectroscopy takes less than a minute, and is
achieved by simply moving a slit and a grism into the beam. A switch
between imaging/spectroscopy typically requires the telescope to be
refocussed.
Camera |
Wavelength Range |
Dispersion (Å/pixel) |
Slit Length |
Detector |
Comments |
WFC low-res spectroscopy |
Z 0.83 - 0.95 µm
Y 0.955 - 1.085 µm
Yn 1.00 - 1.05 µm
J 1.16 - 1.33 µm
H 1.48 - 1.78 µm
K 1.95 - 2.36 µm
|
1.7 in Z
1.9 in Y
2.1 in Yn
2.5 in J
3.1 in H
4.1 in K
|
4 arcmin |
HAWAII 1kx1k HgCdTe |
R=2500 with 0.6arcsec slit |
HRC med-res spectroscopy |
Pa-beta 1.26-1.34 µm
1.57-1.67 µm
Br-gamma 2.07-2.20 µm |
0.8 1.0 1.4 |
80 arcsec |
HAWAII 1kx1k HgCdTe |
R~5500 with 0.2arcsec slit
|
Notes:
For WFC spectroscopy the exact range in the K band depends
on the choice of K filter (K',Ks,K).
For WFC spectroscopy the J-band range is sampled in two orders which
both are sampled on the detector simultaneously (see example
image).
For WFC spectroscopy the Z-band range (filter Yn #222) is sampled in two orders which
both are sampled on the detector simultaneously (see example
image).
Available slits
For WFC spectroscopy we have a dedicated big slit (approx 0.6 arcsec wide).
For HRC spectroscopy we have a dedicated small slit (approx 0.2 arcsec wide),
but in principle the big slit can be used as well.
There is no wide slit available, but for flux standards a wide pinhole can be used.
Calibration unit
The calibration unit is mounted in the telescope baffle,
and can only be used with the mirror cover closed. This causes
a 1-minute overhead to open/close the mirror. Note that the Argon lamp
takes a while to switch on (please exercise it in the afternoon).
Arc maps and sky lines
Typical exposure
times for the calibration lamps (halogen,Argon,Xenon)
System efficiency, expected count rates, etc.
Science-grade array
Exposure-time calculator
Run preparation
Create NOT TCS target catalogue
(make sure to include your standards as well).
ING source visibility plot form
NOTCAM script creation:
to create batch files for dithering, etc. Note that nodding along the slit corresponds
to a movement in the X direction.
User manuals
NOTCAM users guide aimed at imaging, but with a lot of
general info about NOTCAM
NOTCam Cookbook (NOTCam observer's manual)
For more info on NOTCAM spectroscopy, see the first-light and commisioning reports below.
For more info on NOTCAM in general, see the NOTCAM main page
Commissioning results
Reports on commissioning, first observations, possible upgrades
Calibration unit
Pictures
Staff routines
How to focus and align slits
Slit shape monitoring
John Telting (jht)
|