Search
Text Size:
Smaller Text Normal Text Larger Text

Facebook page

Twitter profile

YouTube channel

NOT CCD #7 test report.

Data collected 99 01 05.

Test circumstances.

LocationIn ALFOSC, on telescope.
Light sourceBlue light in filter wheel.

Summary of Results

Bias level390ADU
Bias std. dev.5.27ADU
Conversion factor1.062electrons/ADU
Readout Noise5.60electrons/pixel
Nonlinearity<0.1%
Shutter error0.025sec

Data

Window used: (896:1152,896:1152) (pixel, of full field, unbinned).

Exposure time (s)Mean Counts 1 (ADU)Mean Counts 2 (ADU)(Photon Noise)2 (ADU2)
2.000002528.012528.122497.16
2.000002534.732534.962469.52
4.000005025.535026.694891.94
8.0000010041.510041.79723.00
12.000015025.415029.014502.1
16.000020056.820058.619404.4
20.000025038.325042.823845.0
24.000030080.130080.729120.0
28.000035064.235066.633464.6
32.000040108.740109.838509.2
36.000045099.945104.542963.8
40.000050144.150147.547345.6
44.000055148.055151.652215.9
48.000060186.560189.656393.6
52.000064929.764930.822383.5
This table as ASCII

Plots

Linearity as count rate GIF PostScript
Photon transfer curve GIF PostScript
BitStat GIF PostScript

Images

A median stacked image of 21 bias frames is available here (2.7 MByte, gzipped).

A median stacked image of 20 short exposures (0.1 sec each) is available here (4 MByte, gzipped). This image may be useful for exploring the blemish structure of CCD#7. There are several traps and a couple of bad columns. The many spots with reduced sensitivity are artefacts of the thinning/coating process.

Back to top Last modified: May 31 2023